Commit d100e50b authored by Tobias Steinhoff's avatar Tobias Steinhoff

Update README.md

parent 25fa5918
......@@ -9,7 +9,8 @@ GO Labview programm was adjusted for the following points:
Hardware settings/changes:
* shortcut equilibrator pump and set pump speed to 0
* pressure was measured in the gas line with the LI6262 when measuuring equilibrator air
* The pressure for the standard gases atteh gas regulators was set to 4 PSI. The pressure for the standard gases in the gas stream when gas was flowing was adjusted to teh pressure measured during equilibrator measurements using the needle valves.
* The pressure for the standard gases atteh gas regulators was set to 4 PSI. The pressure for the standard gases in the gas stream when gas was flowing was adjusted
* to the pressure measured during equilibrator measurements using the needle valves (+- 10 mbar).
*
## test runs
......@@ -17,5 +18,5 @@ the instrument was connected in line with a LI6262.
<img src="figures/LI7815-GO-setup.png" width="600">
The pressure of the gas standras was set to 4 PSI.
Markdown is supported
0% or
You are about to add 0 people to the discussion. Proceed with caution.
Finish editing this message first!
Please register or to comment